SKU
HI3001

Flow-thru conductivity electrode - HI3001

244,00 €
Na zalogi
This four-ring flow-thru probe measures conductivity in industrial applications with a platinum sensor and NTC sensor for Automatic Temperature Compensation. With it's durable design, you can count on this electrode to stand up to your process environment.
  • Four-ring probe measures conductivity with platinum sensors.
  • Houses an NTC sensor for Automatic Temperature Compensation.
  • Built for everyday, demanding use in industrial applications.

Four-Ring Probe

The HI3001 is a potentiometric sensor, utilizing four-ring technology to achieve accurate EC measurements. The four-ring technology allows for a wide range of measurement with a single probe as compared to amperometric two-pole probes that have a limited range and can suffer from polarization effects. The platinum sensors are protected by a PEI cover that can easily be removed for quick maintenance.

Internal Temperature Sensor

The integrated NTC temperature sensor of the HI3001 probe is crucial for accurate EC measurements. Since the temperature of a solution can drastically change the amount of conductance, obtaining a fast and stable temperature measurement allows for an accurately compensated EC reading.

Threaded Connection

The ½” front NPT threaded connection of the HI3001 allows for flow-thru mounting, while the ¾” back threads allow for submersion or pipe mounting without the need for additional hardware.

PVDF Body

Polyvinylidene fluoride (PVDF) is a plastic that is resistant to most chemicals and solvents. It has high abrasion resistance, mechanical strength, and resistance to ultraviolet and nuclear radiation. PVDF is also resistant to fungal growth.

Multiple Cable Lengths

The HI3001D models feature a DIN connector for use with Hanna’s HI99XX series of wall-mounted controllers, along with varying lengths of cable. HI3001D comes with a DIN connector and 3 meters (9.9’) of cable attached. HI3001D/5 comes with a DIN connector and 5 meters (16.4’) of cable attached. HI3001D/10 comes with a DIN connector and 10 meters (32.8’) of cable attached.

PEI Protective Cover

The protective cover is made of PEI and can be removed for quick maintenance. These probes can withstand temperatures up to 80°C (176°F) and 6 bars (87 psi) of pressure.

 

Specifications

 

Body Material

PEI, PVDF

Temperature Compensation Range

Automatic, 0 to 60°C with NTC sensor

Max. pressure

6 bar (87 psi)

Overall Length

89mm(3.5”)

Cable

3 m (9.9’), 5 m (16.4’) and 10 m (32.8’) based on selection

Connection Type

N/A (HI3001), DIN (HI3001D models)