Dissolved oxygen controller with Extended Range and Analog Output - HI8410
- Fail Safe Alarm System protects these meters against the pitfalls of process control.
- Low maintenance Galvanic probe.
Fail Safe Alarm Syste
Hanna’s exclusive Fail Safe Alarm System protects against problems caused by power supply failure or signal interruption, which are typical risks in industrial environments. An alarm condition is indicated by a red LED, located directly on the front panel of the controller.
Automatic Temperature Compensation
The Dissolved Oxygen readings are automatically compensated for the temperature effects on the oxygen solubility and membrane permeability.
Salinty Compensation
The salinity compensation feature allows the determination of Dissolved Oxygen even in salty waters.
Simple to Use
The instruments are designed with a standard DIN panel mount with membrane keypads and large LCD with backlight on the front, and provide a series of auto-diagnostic functions.
Analog Output
To enhance troubleshooting and the ability to record data while monitoring, simply attach a recording device to the instrument’s 4 to 20 mA output contacts, conveniently located on the front panel, to obtain a copy of the results on demand.
Diagnostic Tests
The HI 8410 controllers are designed with built-in diagnostic functions to enable the user to check and troubleshoot the instrument. The checks performed are through the front panel keys and can be used to isolate the cause of malfunction.
Specifications
Dissolved Oxygen Range |
0.0 to 50.0 mg/L (ppm) O₂; 0 to 600 % O |
Measurement Type galvanic |
HI76410/4 with 4 m (13.1’) cable or HI76410/10 with 10 m (32.8’) cable (not included) |
Resolution |
0.1 mg/L (ppm) or 1% (O₂) |
Accuracy |
±1% of reading (O₂) |
Temperature Compensation |
automatic, from -5 to 50°C (23 to 122 °F) |
Salinity Compensation |
0 to 51 g/L (resolution 1 g/L) |
Calibration |
manual, one point, in saturated air |
Dosing Relay |
1, isolated, 2A, max. 240V, resistive load, 1,000,000 strokes |
Set Point |
1 to 600 % O2; 0.1to 50.0 mg/L (mg/L (ppm) O₂ |
Hysteresis Band |
0.5 to 2.4 mg/L (ppm) O₂ |
Alarm Relay Range |
1.0 to 5.0 mg/L (ppm) O₂ |
Dosing Control |
OFF/AUTO/ON with selection switch |
Over-Dosing Control |
adjustable, from 5 min to 60 min with knob or disable by wire strap - on rear panel |
Backlight |
continuous on |
Enclosure flame |
retardant ABS body and front panel; transparent splash-proof front cover Panel Cutout 141 x 69 mm (5.6 x 2.7’’) |
Recorder Output |
0 to 20 mA or 4 to 20 mA (isolated) |
Power Supply |
115 VAC ±10%; 50/60 Hz |
Environment |
10 to 50°C (14 to 122°F); RH max 95% non-conndensing |
Weight |
1 kg (2.2 lb.) |
Ordering information |
The HI8410 is supplied with mounting brackets and instructions. Probes (not included): HI76410/4: Galvanic DO probe (fixed) with internal temperature sensor, DIN connector and 4 m (13.1’) cable HI76410/10: Galvanic DO probe (fixed) with internal temperature sensor, DIN connector and 10 m (32.8’) cable |