SKU
HI4107
Chloride Combination Ion Selective Electrode (ISE) - HI4107
907,00 €
Na zalogi
The HI4107 is a solid state, chloride combination ion selective electrode (ISE) for the determination of chloride (Cl-) in solution. The silver chloride sensing pellet produces a potential change due to the chloride ion exchange between the inorganic pellet membrane and the sample. The internal sensing elements are housed within a durable polyetherimide (PEI) body. The HI4107 is ideal for a variety of applications in food and beverages, agriculture, and water treatment.
- Solid state sensor
- Skirted cone geometry
- Detection from 1.8 to 35,500 mg/L Cl-
The HI4107 chloride ion selective electrode is a complete potentiometric cell that contains a silver/silver chloride (Ag/AgCl) reference housed within a PEI plastic body with a potassium nitrate electrolyte. The internal elements are isolated from the sample by a compressed silver chloride sensing membrane.
Conversion Table for Cl-
Multiply | |
moles/L (M) to ppm (mg/L) |
3.500x104 |
ppm (mg/L) to M (moles/L) |
2.857x10-5 |
Features at-a-glance
Fixed Sensing Modules
- Combination solid-state ISEs are electrodes that detect free ions in a solution. The highly insoluble, compressed sensing pellet is attached to the electrode by a fixed sensing module for ease of use.
Durable PEI Body
- The probe body is composed of durable polyetherimide (PEI) plastic.
BNC Connection
- The HI4107 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.
Theory of Operation
A solid-state sensor electrode develops a voltage due to the ion exchange that occurs between the solution and the inorganic membrane. An equilibrium occurs due to the highly limited solubility of the membrane material in the sample. When the ionic strength of the solution is fixed by the addition of ISA, the voltage is proportional to the concentration of free ions in solution.Specifications
Specification Name | Detail |
---|---|
SKU | HI4107 |
Type | solid state; combination |
Optimum pH Range | 2 to 11 |
Temperature Range | 0 to 80°C |
Diameter | 12 mm |
Length | 120 mm |
Body Material | PEI |
Cable | 1 m coaxial |
Common Applications | determination of free chloride ions in emulsified food products, beverages, plants, soils and as an indicator for titration |
Connection | BNC |